ANALYSIS OF OXYGEN CONCENTRATION VALUES IN PORTABLE OXYGEN CONCENTRATOR DEVICES BY USING SILICA GEL AS FILTRATION MATERIAL

Authors

  • Syahrul Ramadhan Ardiansyah Universitas PGRI Adi Buana Surabaya
  • Akbar Sujiwa Universitas PGRI Adi Buana

DOI:

https://doi.org/10.36456/best.vol5.no1.7891

Keywords:

concentra, concentrator, filter, silica

Abstract

Oxygen concentrators as a means of separating pure oxygen from free air became an item that was quite sought after when the pandemic case hit in the past 2020 to 2021. This device works by taking free air around us and then processing it by passing the air through a filter so that oxygen is separated from other gases, most of which are nitrogen. However, not all oxygen generator devices are able to filter oxygen to the maximum, this depends on the filtration material used. Most portable oxygen generator devices use synthetic zeolite as a filtration material, but have not maximally filtered oxygen, therefore in this study researchers examined the effect of using silica gel when used as an oxygen filtration medium in the same oxygen concentrator device. In the results of this study, it was found that there was an increase in the oxygen concentration from the standard media device using synthetic zeolite compared to using silica gel material, the increase in oxygen concentration that occurred increased by around 3.3%.

References

Sujiwa, A., & Raharjo, I. P. (2021). Design and Construction of Automatic Portable Disinfectant Button to Prevent the Spread of The Covid-19 Virus. BEST: Journal of Applied Electrical, Science, & Technology, 3(2), 1-4.

Bordes, J., Erwan d’Aranda, Savoie, P. H., Montcriol, A., Goutorbe, P., & Kaiser, E. (2014). FiO2 delivered by a turbine portable ventilator with an oxygen concentrator in an Austere environment. The Journal of Emergency Medicine, 47(3), 306–312.

Moll, J. R. onald., Vieira, J. E. dso., Gozzani, J. L. auz., & Mathias, L. A. S. ilv. T. (2014). Oxygen concentrators performance with nitrous oxide at 50:50 volume. Brazilian Journal of Anesthesiology (Elsevier), 64(3), 164–168.

Graziano, L., Asciutti, D., Savi, D., Rivolta, M., Turinese, I., Schiavetto, S., Perelli, T., Bertasi, S., Valente, D., & Palange, P. (2019). P381 Efficacy of a portable oxygen concentrator in the promotion of physical activity and the quality of life in a group of patients with cystic fibrosis: pilot study. Journal of Cystic Fibrosis, 18, S165.

Daffa Nabilah Kartika. (2020). “Analisis Pressure Swing Adsorption pada Material Adsorbent untuk Aplikasi Oxygen Concentrator.” Skripsi. Program Sarjana Universitas Padjadjaran (Unpad). Sumedang.

Nurrahmawati, A., Harmadi, H., & Okta Biolita, N. (2018). Rancang Bangun Alat Ukur Konsentrasi Oksigen yang Dihasilkan oleh Fotobioreaktor Mikroalga Chlorella Vulgaris. Jurnal Otomasi Kontrol Dan Instrumentasi, 10(1), 49. https://doi.org/10.5614/joki.2018.10.1.5

Widiatmoko, A., Gede, I. D., Wisana, H., & Rahmawati, T. (2019). Rancang Bangun Pengukur Konsentrasi Oksigen Pada Alat Bubble CPAP. 8, 182–188.

Zhang, Q., Liu, Y., Li, Z., Xiao, P., Liu, W., Yang, X., Fu, Y., Zhao, C., Yang, R. T., & Webley, P. A. (2021). Experimental study on oxygen concentrator with wide product flow rate range: individual parametric effect and process improvement strategy. Separation and Purification Technology, 274(January), 118918. https://doi.org/10.1016/j.seppur.2021.118918

A. F. Al-Shawabkeh, N. Al-Najdawi, and A. N. Olimat, “High purity oxygen production by pressure vacuum swing adsorption using natural zeolite,” Results in Engineering, vol. 18, p. 101119, Jun. 2023, doi: 10.1016/J.RINENG.2023.101119.

M. Meriatna, L. Maulinda, M. Khalil, and Z. Zulmiardi, “PENGARUH TEMPERATUR PENGERINGAN DAN KONSENTRASI ASAM SITRAT PADA PEMBUATAN SILIKA GEL DARI SEKAM PADI,” Jurnal Teknologi Kimia Unimal, vol. 4, no. 1, pp. 78–88, Nov. 2017, Accessed: Aug. 06, 2023. [Online]. Available: https://ojs.unimal.ac.id/jtk/article/view/65

S. Sugiarti, C. Charlena, and N. A. Aflakhah, “Zeolit Sintetis Terfungsionalisasi 3-(Trimetoksisilil)-1-Propantiol sebagai Adsorben Kation Cu(II) dan Biru Metilena,” Jurnal Kimia VALENSI, vol. 3, no. 1, May 2017, doi: 10.15408/JKV.V0I0.5144.

Downloads

Published

31-03-2023

Issue

Section

Contents of the Journal

How to Cite

Ardiansyah, Syahrul Ramadhan, and Akbar Sujiwa. “ANALYSIS OF OXYGEN CONCENTRATION VALUES IN PORTABLE OXYGEN CONCENTRATOR DEVICES BY USING SILICA GEL AS FILTRATION MATERIAL”. Best : Journal of Applied Electrical, Science and Technology, vol. 5, no. 1, Mar. 2023, pp. 11-14, https://doi.org/10.36456/best.vol5.no1.7891.